Application of the Hartmann formula for calibration of prism spectral devices in a wide range

Abstract

Application of the Hartmann formula for calibration of prism spectral devices in a wide range

Shirokikh T.V.

Incoming article date: 02.03.2024

Prism spectral devices have a spectrum-variable dispersion and therefore a nonlinear relationship between the wavelength and the position of spectral lines on the focal surface of a spectral monochromator device, which makes it difficult to calibrate such devices in terms of wavelengths and dispersion.Most often, the well-known Hartmat formula is used for graduation. However, the accuracy of its calculation is satisfactory only in a limited spectral range, and therefore the calculation is carried out on overlapping areas no wider than 200 nm with averaging in overlapping zones. Averaging gives a calibration curve (and, accordingly, a dispersion curve) with gaps at the joints. In this paper, we consider the possibility of using a single calibration for the entire spectrum region, providing smooth, continuous calibration curves. The best result is obtained by using the Hartmann formula for the entire spectral range, after which a set of deviations of the calibration points is determined. This set is interpolated by a polynomial of an arbitrarily chosen order n using the least squares method.

Keywords: monochromator, spectrum, calibration, dispersion, Hartmann formula, interpolation, polynomial